Skip to main content

Advertisement

Table 7 Stresses on the hard bone and soft bone around the implant area in consolidated and non-consolidated in group I and group II models (Mpa)

From: Evaluating the effects of consolidation on intrusion and retraction using temporary anchorage devices—a FEM study

   Max von Mises stress in consolidation Max von Mises stress in non-consolidation Max principal stress in consolidation Max principal stress in non-consolidation
Hard Soft Hard Soft Hard Soft Hard Soft
Group I 7 mm 29.04 0.99 29.1 0.99 32.85 1.15 32.9 1.15
10 mm 25.9 0.6 26 0.6 29.3 0.72 29.3 0.72
13 mm 22.35 0.5 22.42 0.5 26.17 0.56 26.26 0.56
Group II 7 mm 35.9 1.24 35.9 1.24 40.9 1.39 41 1.39
10 mm 25.67 0.6 25.75 0.6 29.35 0.7 29.45 0.7
13 mm 23.74 0.049 23.59 0.5 31.04 0.51 30.85 0.51
  1. Group I: Von Mises stresses and principle stresses are more in both consolidations and non-consolidation in the hard and soft bone at 7-mm implant height
  2. Group II: Von Mises stresses and principle stresses are more in both consolidations and non-consolidation in the hard and soft bone at 7-mm implant height