Skip to main content

Table 2 Comparison of forces and moments generated by direct printed aligners of the 5 tested groups on the UL1

From: Forces and moments generated by 3D direct printed clear aligners of varying labial and lingual thicknesses during lingual movement of maxillary central incisor: an in vitro study

Force

Group 1

Group 2

Group 3

Group 4

Group 5

P valuea

Median

IQR

Median

IQR

Median

IQR

Median

IQR

Median

IQR

Fx

1.44

1.22–1.80

1.00

0.78–1.17

1.01

0.86–1.15

1.04

0.89–1.21

0.90

0.76–1.05

< 0.001

Fy

− 0.27

− 0.64 to 0.03

− 0.51

− 0.75 to − 0.26

− 0.24

− 0.55–0.14

− 0.25

− 0.46 to 0.00

− 0.23

− 0.48 to 0.08

< 0.001

Fz

0.12

− 0.10 to 0.25

0.11

0.03–0.21

0.14

0.01–0.27

0.19

− 0.04–0.33

− 0.03

− 0.17 to 0.14

< 0.001

Mx

5.97

2.29–8.08

0.75

− 0.85 to 3.48

4.06

0.86–5.97

2.02

0.08–4.06

0.11

− 2.74 to 2.56

< 0.001

My

8.96

7.69–11.33

4.95

3.33–6.11

6.10

4.55–8.55

7.35

6.40–8.30

6.16

4.69–7.58

< 0.001

Mz

− 20.74

− 25.71 to − 18.19

− 11.26

− 12.81 to − 9.05

− 14.98

− 16.28 to − 13.17

− 12.67

− 16.57 to − 10.78

− 11.89

− 15.00 to − 9.28

< 0.001

  1. P valuea—overall p value for the model