Evaluation of shear bond strength of orthodontic brackets using trans-illumination technique with different curing profiles of LED light-curing unit in posterior teeth

Abstract

Background

Although using light-cured composites for bonding orthodontic brackets has become increasingly popular, curing light cannot penetrate the metallic bulk of brackets and polymerization of composites is limited to the edges. Limited access and poor direct sight may be a problem in the posterior teeth. Meanwhile, effectiveness of the trans-illumination technique is questionable due to increased bucco-lingual thickness of the posterior teeth. Light-emitting diode (LED) light-curing units cause less temperature rise and lower risk to the pulpal tissue. The purpose of this study was to evaluate the clinical effectiveness of trans-illumination technique in bonding metallic brackets to premolars, using different light intensities and curing times of an LED light-curing unit.

Methods

Sixty premolars were randomly divided into six groups. Bonding of brackets was done with 40- and 80-s light curing from the buccal or lingual aspect with different intensities. Shear bond strengths of brackets were measured using a universal testing machine. Data were analyzed by one-way analysis of variance test and Duncan's post hoc test.

Results

The highest shear bond belonged to group 2 (high intensity, 40 s, buccal) and the lowest belonged to group 3 (low intensity, 40 s, lingual). Bond strength means in control groups were significantly higher than those in experimental groups.

Conclusions

In all experimental groups except group 6 (80 s, high intensity, lingual), shear bond strength was below the clinically accepted values. In clinical limitations where light curing from the same side of the bracket is not possible, doubling the curing time and increasing the light intensity during trans-illumination are recommended for achieving acceptable bond strengths.

Background

Acid etch technique was first introduced by Buonocore in 1955[1], and since then, it has become increasingly popular among dentists due to its numerous advantages. It has been introduced by Newman[2] in the field of orthodontics and used for direct bonding of brackets for many years.

Using self-polymerizing composites for bonding brackets needs significant time for mixing, so it is a time-consuming procedure when multiple mixes are required. The polymerizing process begins immediately after mixing; therefore, limited time is available for positioning the brackets. Another disadvantage of this method is the possibility of air bubble lockup into the adhesive during mixing that would reduce the bond strength of orthodontic brackets. However, using light-polymerizing composites provides extended working time for prompt bracket positioning and easier residue removal[3]. Moreover, it has been demonstrated that the initial bond strength of brackets would be higher by using light-polymerizing composites rather than self-polymerizing materials[4].

Since curing light cannot penetrate the metallic bulk, polymerization of composites under metallic brackets is limited to the edges of the bracket base. So it may result in incomplete polymerization and diminished bond strength[3]. Some researchers have suggested that curing the composite layer under metallic brackets can be performed from mesial and distal aspects for 20 s each[5]. However, limited access to and poor direct sight of posterior segments may cause some difficulties especially when curing the composite under the bracket from the distal aspect. Therefore, trans-illumination technique has been proposed to cure the composite under metallic brackets by Tavas and Watts in 1979[6]. In this technique, light is emitted from the opposite side of the tooth and passes through bucco-lingual thickness toward the composite under the metallic bracket. Although little information can be found in orthodontic literature to confirm its usefulness, there is a general agreement on increasing the curing time while using this technique[7]. King et al.[8] tripled the trans-illumination curing time and found proper shear bond strength values regardless of the bucco-lingual thickness of the teeth ranging from 3.4 to 7 mm. Oesterle and Shellhart[7] used human maxillary incisors and found out that the bond strength of 50-s curing from the lingual aspect was so close to the control group and there was no significant difference in adhesive remaining index between control and experimental groups.

There are various types of light-curing units with different spectral profiles and light intensities. While low-intensity light may lead to inadequate depth of cure and insufficient bond strength, high-intensity light might cause excessive heat during irradiation[911]. When using trans-illumination technique, the concerns over pulp chamber temperature rise become more important because the light passes through the bucco-lingual thickness of the tooth including the pulp chamber. Many studies have been done to compare pulp chamber temperature rise during polymerization while using different types of light-curing units. Yazici et al.[12] and Haiduc et al.[13] reported that using light-emitting diode (LED) light-curing unit results in significantly lower temperature rise in comparison with conventional halogen units.

As in some cases, it is proposed to have a trans-illumination for bonding of posterior brackets due to limited access from the buccal side to their distal aspect and also for being assured from complete polymerization of bonding materials under the metallic brackets. The purpose of this study was to evaluate the effectiveness of trans-illumination technique in bonding metallic brackets to premolars while using different curing profiles of LED light-curing unit.

Methods

Based on the results of Oesterle and Shellhart's study[7] and the following formula, the sample size was found to be 10:

n = N δ 2 N - 1 D + δ 2 , D = B 2 4

where n is the sample size, N is the population size, δ2 is the variance, and B is the error bound. To estimate δ2, we used an elementary sample with α = 0.05. The power of the study was measured to be 80%. Sixty human premolars that were recently extracted for orthodontic purposes were collected. The criteria for tooth selection were intact buccal enamel, no pretreatment of chemical agents (such as derivatives of peroxide, acid, or alcohol), no cracks from forceps, no caries, and no restorations.

The teeth were sterilized in buffered formalin as Lee et al. suggested in 2007[14]. One week prior to use, the teeth were placed in isotonic normal saline (0.90% w/v of NaCl, 300 mosM/L) in order to avoid any possible effect of the remaining solution on the bonding process. The storage media was changed every day to avoid bacterial growth. The teeth were embedded in auto-polymerizing polymethyl methacrylate. A mounting jig was used to align the facial surface of the teeth to be perpendicular to the bottom of the mold and its labial surface parallel to the force during the shear bond strength (SBS) test.

Before bonding, the teeth were randomly divided into six groups each containing ten teeth. Specimen preparation was done exactly as was instructed by the manufacturer. The labial surfaces of the teeth were polished using non-fluoride pumice and then rinsed with water and subsequently dried with moisture-free air. The buccal enamel was etched with a 37% phosphoric acid (UltraEtch, Ultradent Products Inc., South Jordan, UT, USA) for 30 s. Afterwards, the etched surface was rinsed for at least 15 s until the etchant was completely removed and then dried with oil and moisture-free air source. A thin uniform coat of Transbond XT primer (3 M Unitek, Monrovia, CA, USA) was applied to the etched surfaces. A small amount of adhesive paste of Transbond XT was applied to the bracket bases. Stainless steel brackets used in this study were Ultratrim Standard Edgewise (Dentaurum, Ispringen, Germany) in all groups. Immediately after placing the adhesive, the brackets were lightly placed on the tooth surface, adjusted to the final position and then pressed firmly. Excess adhesive material was gently removed from around the bracket base without disturbing it.

The adhesive was cured with High Power (800 mW/cm2) and Low Power (650 mW/cm2) programs of Bluephase C8 (Ivoclar, Vivadent, Schaan, Liechtenstein) LED-curing unit in all groups:

  • Group 1 (control): 40 s light curing with the Low Power program from the buccal aspect (10 s for each mesial, distal, occlusal, and gingival aspect).

  • Group 2 (control): 40 s light curing with the High Power program from the buccal aspect (10 s for each mesial, distal, occlusal, and gingival aspect).

  • Group 3 (experimental): 40 s light curing with the Low Power program from the lingual aspect (light cure tip was placed as close as possible to the lingual surface perpendicular to occluso-gingival axis of the tooth).

  • Group 4 (experimental): 80 s light curing with the Low Power program from the lingual aspect (light cure tip was placed as close as possible to the lingual surface perpendicular to occluso-gingival axis of the tooth).

  • Group 5 (experimental): 40 s light curing with the High Power program from the lingual aspect (light cure tip was placed as close as possible to the lingual surface perpendicular to occluso-gingival axis of the tooth).

  • Group 6 (experimental): 80 s light curing with the High Power program from the lingual aspect (light cure tip was placed as close as possible to the lingual surface perpendicular to occluso-gingival axis of the tooth).

After bonding, specimens were stored individually in a normal saline solution at 37°C in a dark environment 24 h prior to testing. The shear bond strength of specimens was measured using a Zwick testing machine (Zwick GmbH & Co, Ulm, Germany) at a crosshead speed of 0.5 mm/min. In order to avoid bias, the bonding and debonding procedures were done by two different operators and the teeth were given codes unrelated to their group numbers. The results of the SBS test were recorded in megapascal.

Statistical analysis

The data was confirmed to be normally distributed using the Kolmogorov-Smirnov test. One-way analysis of variance (ANOVA) and Duncan's post hoc tests were used to compare SBS values among groups using SPSS 11.5 software.

Results

The one-way ANOVA test showed that there was a significant difference among shear bond strength amounts in groups (F(5,10) = 719.394, p < 0.05). Also, as it is shown in Table 1, Duncan's post hoc tests revealed that there was a significant difference between every two groups except for groups 1 and 2 (control group).

Table 1 Duncan's post hoc test

Mean values of SBS are shown in Table 2. The highest mean bond strength was seen in group 2 (40-s curing from the buccal aspect with the intensity of 800 mW/cm2) and the lowest value was in group 3 (40-s curing from the lingual aspect with the intensity of 650 mW/cm2).

Table 2 Shear bond strength values in all six groups

SBS was not significantly different between two control groups; in both groups, the SBS values were above the clinically accepted values according to Reynolds' study[15], and values were significantly higher than experimental groups.

Among four experimental groups, the only group with sufficient SBS values for orthodontic bonding was group 6 (80-s curing from the lingual aspect with the intensity of 800 mW/cm2).

Discussion

Tavas and Watts[6] first introduced trans-illumination technique in 1979, which suggested curing composite adhesive under metallic brackets from the opposite side of the tooth. Many studies have been done on this technique mostly on the anterior teeth, and the effectiveness of this technique has been proved in the anterior region. King et al.[8] studied trans-illumination technique using bovine teeth and lingual brackets. They tripled the curing time and found proper shear bond strength values when using trans-illumination regardless of the bucco-lingual thickness of the teeth ranging from 3.4 to 7 mm. Oesterle and Shellhart[7] used human maxillary incisors and compared shear bond strength when curing from the buccal (as control groups) or lingual (trans-illumination technique) aspect. They found bond strength values very close to control groups with 50-s curing from the lingual aspect and no statistical difference in adhesive remaining index between control and experimental groups.

Although in posterior segments, limited access and poor direct sight cause some difficulties in curing the adhesive from the same side of the bracket (especially when it is needed to be cured from the distal or gingival aspect), there is little in orthodontic literature about the effectiveness of trans-illumination technique in posterior segments. Regardless of the conclusion of each study, there is a general agreement on using higher light energy (light intensity or curing time) while directing the light through the bucco-lingual thickness of the tooth[7]. However, possible damaging effects of excessive heat generated during polymerization have been warned[911]. Previous studies have shown increase in temperature ranging from 1.5°C to more than 4°C in the pulp chamber of extracted teeth[9, 10] which were assumed to be caused by both exothermic reaction and energy absorbed during irradiation[1620]. The critical temperature for making irreversible damage to the pulpal tissue is 42°C to 42.5°C[21, 22]. As Yazici et al.[12] and Haiduc et al.[13] demonstrated in their studies, using LED units can cause significantly lower temperature rise in the pulp chamber in comparison with halogen units. Consequently, in our study, a polywave LED unit, Bluephase C8 (Ivoclar, Vivadent, Schaan, Liechtenstein), was used in bonding procedures in order to make minimum temperature changes in the pulp chamber.

As it is shown in Table 2, the highest shear bond strength was in group 2 (40-s curing from the buccal aspect with the intensity of 800 mW/cm2) and the lowest value was in group 3 (40-s curing from the lingual aspect with the intensity of 650 mW/cm2).

Increase in light intensity from 650 to 800 mW/cm2 between the control groups did not result in any significant increase in SBS values. Among the four experimental groups, the only group with sufficient SBS values for orthodontic bonding was group 6 (80-s curing from the lingual aspect with the intensity of 800 mW/cm2). In group 4, increasing the curing time in comparison with group 3 resulted in a significant increase in bond strength, but according to Reynolds' study[15], the values were not acceptable for orthodontic treatment. Also, in group 5, increasing the intensity (without increasing curing time) in comparison with group 3 resulted in higher bond strengths, but still the values were not acceptable. This leads to the deduction that doubling the light curing time and increasing the light intensity to 800 mW/cm2[2] have to be done simultaneously in order to achieve acceptable bond strength values when using trans-illumination technique.

In another similar study on human premolars using metallic and ceramic orthodontic brackets with different curing times of a halogen unit[23], SBS in control groups (cured from the buccal aspect) was significantly higher than that in experimental groups (cured from the lingual aspect), and in one of their experimental groups with 40-s curing from the lingual aspect, the SBS values were lower than the values accepted for clinical orthodontic treatment as they were in our study. On the other hand, in the other experimental groups of their study, SBS values were clinically acceptable. This difference between the results can be the consequence of the difference in light-curing units used in each study.

Conclusions

According to the results of this study, using trans-illumination technique in bonding metallic brackets to the premolar teeth (in comparison with light curing from the same side of the bracket) resulted in significantly lower bond strengths. In clinical limitations where light curing from the same side of the bracket is not possible, doubling the curing time and increasing the light intensity to 800 mW/cm2 during trans-illumination with the LED light-curing unit can be done for attaining acceptable bond strength with minimum risk of overheat to the pulpal tissue.

References

  1. Buonocore MG: A simple method of increasing the adhesion of acrylic filling material to enamel surfaces. J Dent Res. 1955, 34: 849–53. 10.1177/00220345550340060801

    Article  PubMed  Google Scholar 

  2. Newman GV: Epoxy adhesives for orthodontic attachments: progress report. Am J Orthod. 1965, 51: 901–12. 10.1016/0002-9416(65)90203-4

    Article  PubMed  Google Scholar 

  3. Sunna S, Rock WP: An ex vivo investigation into the bond strength of orthodontic brackets and adhesive systems. Br J Orthod. 1999, 26: 47–50. 10.1093/ortho/26.1.47

    Article  PubMed  Google Scholar 

  4. Wendl B, Droschl H: A comparative in vitro study of the strength of directly bonded brackets using different curing techniques. Eur J Orthod. 2004, 26: 535–44. 10.1093/ejo/26.5.535

    Article  PubMed  Google Scholar 

  5. Oesterle LJ, Messersmith ML, Devine SM, Ness CF: Light and setting times of visible-light-cured orthodontic adhesives. J Clin Orthod. 1995, 29: 31–6.

    PubMed  Google Scholar 

  6. Tavas A, Watts DC: Bonding of orthodontic brackets by trans-illumination of a light-activated composite: an in vitro study. Br J Orthod. 1979, 6: 207–8.

    Article  PubMed  Google Scholar 

  7. Oesterle LJ, Shellhart WC: Bracket bond strength with trans-illumination of a light-activated orthodontic adhesive. Angle Orthod. 2001, 71: 307–11.

    PubMed  Google Scholar 

  8. King L, Smith RT, Wendt SL, Behrents RG: Bond strengths of lingual orthodontic brackets bonded with light-cured composite resins cured by trans illumination. Am J Orthod Dentofacial Orthop. 1987, 91: 312–5. 10.1016/0889-5406(87)90172-7

    Article  PubMed  Google Scholar 

  9. Thompson HF, Gomez HF, Puckett AD: Pulpal temperature changes after exposure to a light curing source. J Dent Res. 1997, 76: 524.

    Google Scholar 

  10. Bennett GE, Blank FJ: Thermal effects of visible-light and chemically cured resins. J Dent Res. 1984, 63: 259.

    Google Scholar 

  11. Hannig M, Bott B: In-vitro pulp chamber temperature rise during composite resin polymerization with various light-curing sources. Dent Mater. 1999, 15: 275–81. 10.1016/S0109-5641(99)00047-0

    Article  PubMed  Google Scholar 

  12. Yazici AR, Muftu A, Kugel G, Perry RD: Comparison of temperature changes in the pulp chamber induced by various light curing units, in vitro. Oper Dent. 2006, 31: 261–5. 10.2341/05-26

    Article  PubMed  Google Scholar 

  13. Haiduc C, Dodenciu D, Sinescu C, Negrutiu M, Draganescu G, Mara V: Pulp chamber temperature variations using three types of light curing units. Eur Cell Mater. 2007, 13: 19–23.

    Google Scholar 

  14. Lee JJ, Nettey Marbell A, Cook A Jr, Pimenta LA, Leonard R, Ritter AV: Using extracted teeth for research: the effect of storage medium and sterilization on dentin bond strengths. J Am Dent Assoc. 2007, 138: 1599–603.

    Article  PubMed  Google Scholar 

  15. Reynolds JR: A review of direct orthodontic bonding. Br J Orthod. 1975, 2: 171–80.

    Google Scholar 

  16. McCabe JF: Cure performance of light activated composites by differential thermal analysis (DTA). Dent Mater J. 1985, 1: 231–4. 10.1016/S0109-5641(85)80048-8

    Article  Google Scholar 

  17. LIoyd CH, Joshi A, McGlynn E: Temperature rises produced by light sources and composites during curing. Dent Mater J. 1986, 2: 170–4. 10.1016/S0109-5641(86)80030-6

    Article  Google Scholar 

  18. Masutani S, Setcos JC, Schnell RJ, Phillips RW: Temperature rise during polymerization of visible light activated resins. Dent Mater J. 1988, 4: 174–8. 10.1016/S0109-5641(88)80059-9

    Article  Google Scholar 

  19. Smail SRJ, Petterson CJW, McLaundie AC, Strang R: In vitro temperature rises during visible-light curing of a lining material and a posterior composite. J Oral Rehabil. 1988, 15: 361–6. 10.1111/j.1365-2842.1988.tb00169.x

    Article  PubMed  Google Scholar 

  20. Hanson EK, Asmussen E: Correlation between depth of cure and temperature rise of a light-activated resin. Scand J Dent Res. 1993, 101: 176–9.

    Google Scholar 

  21. Zach L, Cohen G: Pulp response to externally applied heat. Oral Surg Oral Med Oral Pathol. 1965, 19: 515–30. 10.1016/0030-4220(65)90015-0

    Article  PubMed  Google Scholar 

  22. Pohto M, Scheinin A: Microscopic observations on living dental pulp II. The effect of thermal irritants on the circulation of the pulp in the lower rat incisor. Acta Odontol Scand. 1958, 16: 303–27. 10.3109/00016355809064115

    Article  Google Scholar 

  23. Heravi F, Bayani S: Changes in shear bond strength of ceramic and stainless steel brackets with different visible light curing times and directions. Aust Orthod J. 2009, 25: 153–7.

    PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Negin Ghaffari.

Additional information

Competing interests

The authors declare that they have no competing interests.

Authors' contributions

The work presented here was carried out in collaboration among all authors. FH designed the methods and experiments and edited the manuscript text. SMM contributed to the design of the study and selection of bonding materials and devices. NG collected and prepared the samples, interpreted the results, and drafted the manuscript. JJ carried out the SBS test and helped in drafting the manuscript. YB bonded the brackets and worked on the data interpretation. All authors read and approved the final manuscript.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 2.0 International License ( https://creativecommons.org/licenses/by/2.0 ), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Reprints and permissions

About this article

Cite this article

Heravi, F., Moazzami, S.M., Ghaffari, N. et al. Evaluation of shear bond strength of orthodontic brackets using trans-illumination technique with different curing profiles of LED light-curing unit in posterior teeth. Prog Orthod. 14, 49 (2013). https://doi.org/10.1186/2196-1042-14-49

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/2196-1042-14-49

Keywords